- Research
- Open Access
Intradermal microbubbles and contrast-enhanced ultrasound (CEUS) is a feasible approach for sentinel lymph node identification in early-stage breast cancer
https://doi.org/10.1186/s12957-015-0736-x
© Xie et al. 2015
- Received: 18 June 2015
- Accepted: 12 November 2015
- Published: 19 November 2015
Abstract
Background
Microbubbles and contrast-enhanced ultrasound (CEUS) is a new technique for locating sentinel lymph node (SLN). The aim of this study is to explore the feasibility of SLNs tracing by CEUS using microbubbles in breast cancer patients and the value of enhancing patterns in diagnosing lymph nodes metastases.
Methods
A clinical trial was registered (trial registration: ChiCTR-DDT-13003778). One hundred and one consecutive consenting patients with breast cancer undergoing SLN biopsy were enrolled. Before the surgery, microbubble was injected periareolarly. Lymphatic drainage pathway was detected by CEUS, and guidewire was deployed to locate the SLN before the operation. Blue dye was also used to help in tracing sentinel lymph node during the operation. The identification rate and the accuracy rate were recorded. Enhancing patterns of lymph nodes were recorded and compared with the pathological diagnosis.
Results
Of the 101 cases, SLNs in 99 cases were successfully identified by at least one tracer, including 98 cases identified by CEUS and 97 cases by blue dye. There was no significant difference between the two methods (P = 0.705). Guidewires were deployed successfully in all 98 cases, and the localized SLNs were all isolated successfully in the following operations. The status of SLNs isolated by CEUS was completely identical to that of the whole axillary lymph node while 7.1 % cases were misdiagnosed as negative by blue dye method. The sensitivity of predicting SLNs metastases by CEUS enhancing pattern was 81.8 %, the specificity was 86.2 %, and the positive and negative predictive values were 75.0 and 90.3 %, respectively.
Conclusions
Microbubbles and CEUS are feasible approaches for SLN identification. The enhancing patterns on CEUS may be helpful to recognize the metastasizing SLNs. This novel method may be a promising technique for the clinical application.
Keywords
- Sentinel lymph node (SLN)
- Sentinel lymph node biopsy (SLNB)
- Microbubbles
- Contrast-enhanced ultrasonography (CEUS)
- Blue dye, SonoVue
- Breast cancer
Background
Axillary lymph node status is an important prognostic factor in patients with breast cancer [1]. Sentinel lymph node (SLN) biopsy has been the standard procedure for axillary staging in early breast cancer with clinical normal nodes [2–5]. Non-sentinel lymph node (NSLN) metastasis is detected in 35–50 % of SLN positive patients [6, 7]. Although a number of studies have investigated models or nomograms for non-sentinel lymph node (NSLN) metastasis prediction, none of them can predict the probability of NSLN metastasis correctly [8–10]. Hence, if any SLN is positive, the standard procedure remains axillary lymph node dissection (ALND).
The definition of sentinel lymph nodes is the initial lymph nodes that receive lymphatic drainage from the primary tumor. Some markers are used to label the sentinel lymph nodes along the lymphatic drainage pathway. The traditional method for sentinel lymph nodes tracing is a dual-labeled technique involving blue dye and radiolabeled colloid [11]. The identification rate of the dual-labeled technique has been shown to be up to 96 %, with false negative rates between 5 and 10 % [12, 13]. Nevertheless, the half-life of the technetium-99m (99mTc) is up to 6 h, while molybdenum-99 (99Mo) decays rapidly, which may restrict the scheduling of radioisotopes production and delivering. Handling and disposal of radioisotopes are also severe challenges for hospital management. All these drawbacks have limited the widespread use of radioisotopes for SLN biopsy. Isosulfan blue, patent blue, and methylene blue are dye tracers that are frequently used. The identification rates of SLN biopsy by blue dye alone range from 66 to 94 %, with false negative rates of 0 to 12 % [14–17]. Considering the acceptable accuracy and absence of radioactive contamination, the blue dye has become the alternative standard control tracer in many studies [18, 19]. Poor tissue penetration is the obvious drawback of blue dye technique. Surgeons have to anatomize the lymphatic vessels cautiously to find the SLN during the procedure.
The constraints of the existing SLN biopsy techniques have led to the development of alternative methods. Indocyanine green fluorescence (ICG) [18–20], superparamagnetic iron oxide nanoparticles (SPIO) [21, 22], and contrast-enhanced ultrasound (CEUS) using microbubbles have been reported as new feasible alternative techniques for the SLN biopsy in breast cancer [23–26]. Sulfur hexafluoride microbubbles is the most popular ultrasound contrast agent at present, which plays a role as an inert gas in vivo. There is no evidence of metabolism of sulfur hexafluoride. Furthermore, diameters of the bubbles range from 2–10 μm (mean 2.5 μm), which are much less than those of the red blood cells (mean 7.2 μm). So the bubbles can easily pass through the blood capillary and lymphatic microvessels, even across the alveolar epithelium. Therefore, the bubbles also can be cleared up by the lung and kidney easily. On the other hand, there are no iodine and proteins in sulfur hexafluoride microbubbles, which prevents a great portion of patients from having allergy. Sulfur hexafluoride microbubbles may be a promising SLN marker considering its safety and reliability. Nevertheless, there are very few published studies of sentinel lymph node biopsy done by CEUS using microbubbles in human beings [27, 28].
The aim of this study is to validate the effectiveness of the identification of SLNs using microbubbles and CEUS in patients with early breast cancer. Also, we attempted to explore the potential value of CEUS in the pre-operative diagnosis of SLN metastases in patients with early breast cancer.
Methods
Patients
Patients (N = 100) and primary tumor (N = 101) characteristics
Baseline characteristics | Number of patients (N = 100) or cases (N = 101) (%) |
---|---|
Patient characteristics | |
Menopause status | |
Premenopausal | 39 (39 %) |
Post-menopause | 61 (61 %) |
BMI | |
<19 | 3 (3) |
19–25 | 59 (59 %) |
≥25 | 38 (38 %) |
Tumor characteristics | |
Orientation | |
Upper outer quadrant | 46 (45.5 %) |
Upper inner quadrant | 27 (26.7 %) |
Lower outer quadrant | 10 (9.9 %) |
Lower inner quadrant | 9 (8.9 %) |
Central quadrant | 9 (8.9 %) |
Tumor size | |
Tis | 4 (4.0 %) |
T1 | 77 (76.2 %) |
T2 | 20 (19.8 %) |
Histology | |
DCIS | 4 (4.0 %) |
Invasive ductal | 81 (80.2 %) |
Invasive lobular | 8 (7.9 %) |
Others | 8 (7.9 %) |
Stagea | |
0 | 4 (4.0 %) |
I | 49 (48.5 %) |
II | 35 (34.7) |
III | 13 (12.9 %) |
Molecular subtype [35] | |
Luminal A | 26 (25.7 %) |
Luminal B | 53 (52.5 %) |
Her-2 positive | 4 (4.0 %) |
Triple negative | 18 (17.8 %) |
Instruments and materials
All sonographic examinations were performed by two experienced sonographer using two uniform GE Logiq E9 scanners (GE Healthcare) equipped with high-frequency linear array probes (ML6–15-D) and contrast pulse sequences (CPS). Microbubbles composed of phospholipid-stabilized membrane and sulfur hexafluoride gas in the inside (SonoVue, Bracco Imaging) were used as ultrasound contrast agent. SonoVue presented as a dry powder and needed to be prepared by 5 mL of normal saline into 8 μl/ml sulfur hexafluoride microbubbles. The ampoule was shaken vigorously more than 30 s every time before injection to ensure a homogeneous suspension of bubbles. Methylene blue (Jumpcan Pharmaceuticals) was used as a blue dye. A 7-cm-long double hook guidewire (20G, LW0077, BARD) was used to locate the sentinel lymph nodes.
CEUS procedure
Before the SLN biopsy, patients received a gray-scale ultrasound examination of the axilla first, so the sonographers had a general idea of the position of lymph nodes. Then 1.5-mL microbubble suspension was intradermally injected into the upper lateral areola. If an incision biopsy was performed in upper outer quadrant previously, we chose the outer edge of the incision as the injection site instead. After injection, subcutaneous lymphatic drainage pathway directing to the axilla could be detected immediately on contrast pulse sequencing in most cases. Moving the probe along the lymphatic channels, the areas of contrast agent accumulation could be detected, which were the enhanced lymph nodes. Usually, the passing time from injection to axillary node mapping was between 5 and 67 s. Contrast agent remained in the SLN for up to 4 min. Massaging the areola for 10 s intensified the image again. If the lymphatic vessel was not detected successfully in a few cases, another injection was given. If no obvious lymphatic filling was obtained or no lymph node developed after two consecutive injections, the case was recorded as a failure. The conventional gray-scale window was alive synchronously to confirm the presence of an architecturally defined SLN. In some cases, an apparently enhanced lymph node supposed to be an SLN on contrast pulse sequencing was difficult to recognize in a gray-scale pattern.
Once identified, contrast-enhanced ultrasound-identified SLNs were localized with double hook guidewire. After insertion, the tails of the guidewire were covered with a sterile dressing. Considering the unavoidable frequent movements of an upper limb, and the discomfort caused by a foreign matter in the axilla, it was more appropriate to perform this procedure just before the SLN biopsy.
SLN biopsy
Immediately after the induction of general anesthesia, 1-mL blue dye was intradermally injected into the periareolar upper outer quadrant region. If an incision biopsy was performed in upper outer quadrant previously, the outer edge of the incision was taken to be the alternative injection site. Five minutes later, standard SLN biopsies were performed by four well-trained surgeons, respectively. All blue dye-containing lymph nodes, guidewire-directing lymph nodes, and any suspicious lymph nodes during the exploration were excised and numbered in accordance with the following instructions: both blue dye- and guidewire-containing nodes (bub+/dye+) were labeled “1”; only blue dye-containing nodes were labeled (bub-/dye+) “2”, “3” for only guidewire-containing nodes (bub+/dye-), and “4” for neither blue dye- nor guidewire-containing nodes (bub-/dye-). All the lymph nodes were sent for histologic analysis. Standard axillary lymph node dissection was performed if any of the nodes proved positive.
Statistics
The identification rate was defined by the proportion of patients with SLNs identified with either method. Data were subjected to χ 2 test or Fisher’s exact test using SPSS statistical software version 17.0 (SPSS Inc. Chicago, IL, USA). The level of significance was set at P < 0.05.
Results
CEUS technical success
Enhanced lymph nodes (single arrows) on contrast-enhanced ultrasound. a Only one lymph node enhanced. b Two sequentially visualized SLNs linked by lymph vessel. c Three ordinally enhancing SLNs in a row
Guidewire-guided SLNs identification. a Excising guidewire-containing SLN (single arrow) along the wire. Another blue dye-containing lymph node (double arrow) was also isolated. b Two SLNs were excised (single arrow). The guidewire-containing SLN was also reported as “blue” during the procedure
SLN detection
Classification of SLN in terms of CEUS and blue dye
Characteristic | Number of cases (%) | Number of SLNs identified (%) | Number of positive SLNs |
---|---|---|---|
CEUS+/blue dye+ | 96 (95.0 %) | 101 (37.3 %) | 40 |
CEUS+/blue dye− | 2 (2.0 %) | 14 (5.2 %) | 1 |
CEUS−/blue dye+ | 1 (1.0 %) | 107 (39.5 %) | 6 |
CEUS−/blue dye− | 2 (2.0 %) | 49 (18.1 %) | 0 |
Total | 101 (100 %) | 271 (100 %) | 47 |
SLN detection rate according to stage and subtype using the CEUS method
Clinic characteristic | CEUS | P | ||
---|---|---|---|---|
Detection cases | Detection rate (%) | |||
Stage | ||||
0–II | 88 | 88 | 100 | <0.001 |
Stage III | 13 | 10 | 76.9 | |
Subtype | ||||
Triple negative | 18 | 16 | 88.9 | 0.046 |
Others | 93 | 92 | 98.9 |
Pattern of CEUS enhancement
Different types of enhancing patterns; double arrows and circles direct lymph nodes; single arrows direct lymph vessels. a Type I enhancement; SLN was obviously and homogeneously enhanced. b Type II enhancement; the enhancement was obvious not homogeneous. c Type III enhancement; SLN was hardly enhanced
Patterns of enhancement and SLNs status
CEUS | P | |||
---|---|---|---|---|
SLN | Type I | Type II | Type III | |
Pos | 6 | 23 | 4 | <0.001 |
Neg | 56 | 7 | 2 |
Discussion
SLNB has become the routine procedure in patients with early breast cancer, and ALND is still a standard treatment for SLN involved case. Although according to the ACOSOG Z0011 and IBCSG 23-01 data, patients with limited SLN metastatic breast cancer treated with breast conservation and systemic therapy may omit ALND [29, 30]. The conclusions still need to be validated by more well-designed trials. So the primary value of SLNB remains consisting in indicating axillary status with minimal trauma. The combination of radioisotopes and blue dye is a popular method for SLN tracing. However, the limitations are not negligible. The diameter of radioactive colloid is usually between 50 and 100 nm, while the endothelial gap of subcutaneous lymph vessel is between 120 and 500 nm; so the radioactive colloid cannot pass through the lymph–vessel freely but can be absorbed by endocytosis. Therefore, many hours are needed to detect SLNs after radioactive colloid is injected. Moreover, radioactive materials handling is another challenge for hospital management. A blue dye may lead to staining of adjacent adipose tissue. In addition, breast tattooing is another problem that limits its application. Anaphylaxis following blue dye injection is also an infrequent but serious risk for patients. In this study, we chose sulfur hexafluoride microbubbles as a contrast agent, which is much smaller than red blood cell; so the agent can readily cross into the SLNs and be easily eliminated. The detecting instrument is a normal ultrasound apparatus with contrast pulse sequences, and it is readily available. Moreover, there are no iodine and proteins in sulfur hexafluoride microbubbles, which prevents patients from allergy. In consideration of the receivable accuracy rate of blue dye guided method [14–17], and its convenience and fast transit, we chose blue dye as the reference substance for SLNB.
The potential application of CEUS and microbubbles in identifying SLNs in a swine melanoma model was first reported in 2004 [31]. Then, a few studies of SLN identification by CEUS and microbubbles in human beings were published. A UK study with 54 patients found 89 % identification rate by CEUS and intradermal microbubbles [24]. This result was repeated in a further 80-patient UK study [25]. In 2013, a larger study of 347 patients also found a low sensitivity of 61 % but a specificity of 100 % [26]. In the above three studies, there was no significant difference between the identification rates by CEUS and standard dual technique. However, another Japan study with 20 patients had a detecting rate of 70 %, much lower than that of γ-probe-guided method of 100 % [32]. Dual-labeled technique was used as the control method in all the above studies, but the patient numbers varied widely and there was also difference in the effectiveness. In addition, there are few reports on the relationship between the pattern of enhancement and lymph node status.
SLNB using blue dye and radioisotope often results in the removal of multiple SLNs. In this study, the identification rate of CEUS was 97.03 %. One hundred fifteen SLNs were excised by CEUS, which was less than blue dye. The positive rates of SLNs identified by CEUS was 35.7 % (41/115), significantly higher than that by blue dye (21.8 %, 46/211). Rates of axillary seroma and infection in patients who had five or more SLNs removed are significantly higher than those with up to four nodes removed [29]. The characteristic advantage of CEUS and microbubbles is the real-time imaging for SLNs. A dynamic process of SLN enhancement from the afferent lymphatic vessel to efferent vessel is visible, and the very first developed lymph nodes are easily identified and located with a guidewire. Therefore, the number of SLNs detected by CEUS was less than that detected by blue dye in our study, but there was no significant difference between the identification rates by CEUS and blue dye. In other words, maybe SLNB with CEUS can adequately stage the axilla with fewer traumas. The false negative rate of SLNB is the most important concern of the surgeons. It was reported that SLNB had an inherent false negative rate in the range of 4–10 % [33, 34]. In our study, all the axillary positive cases were diagnosed correctly by CEUS except for three undeveloped cases, which underwent ALND consequently and were proved axillary positive pathologically. Therefore, no patient delayed her treatment. There was no unnecessary ALND in the consequence of CEUS, either. However, one case was misdiagnosed negative by blue dye, which may lose the opportunity of ALND without CEUS.
The use of microbubbles and CEUS may also be of value in the diagnosis of involved SLN. The normal direction of internal drainage of lymph nodes is from cortex to cortex, which appeared to be enhanced from peripheral region to central zone in CEUS. SLNs that contained US-detectable metastases demonstrated areas did not enhance as a result of tumor displacement or destruction of the normal tissue. Goldberg had reported that the sensitivity and specificity of CEUS in predicting SLNs metastases in a swine model were 95 and 63 %, respectively, with an overall accuracy of 86 % [31]. In another 20-case study, inhomogeneous enhancement was also observed in 2 metastasized cases. Peripherally increased echogenicity or longer enhancing period was observed because of the replacement of the inner portion of the LN by metastatic lesions [32]. However, there are few reports about the value of the CEUS in predicting lymph node metastases. In our study, we summarized the enhancement pattern into three types and obtained an accuracy rate of 85.2 %, compared with the pathology result. Although the result is not very satisfactory, it is successfully correlated with the intra-nodal filling defects with tumor involved. This may be a valuable addition to the technique.
Conclusions
Microbubbles and CEUS is a feasible approach for SLN identification. The enhancing patterns on CEUS may be helpful to recognize the metastasizing SLNs. This novel method may be a promising technique for the clinical application. Larger patient data and multicenter cohort trials are required and long-term follow-up data on the safety and therapeutic effect are still needed.
Ethics approval and consent to participate
The study was approved by the institutional ethics committee of Peking University People’s Hospital (IRB approval number is 2012045). Informed consent was obtained from all enrolled patients under the ‘Ethics, consent and permissions’ heading (The version number is modified revision 2012.12.18).
Consent for publication
Informed consent for publishing the individual patient data was obtained from all the participants.
Declarations
Acknowledgements
This work was supported by the Foundation for Scientific Research and Development of Peking University People’s Hospital (RDC2012-13).
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
Authors’ Affiliations
References
- Fisher B, Bauer M, Wickerham DL, Redmond CK, Fisher ER, Cruz AB, et al. Relation of number of positive axillary nodes to the prognosis of patients with primary breast cancer. An NSABP update Cancer. 1983;52:1551–7.PubMedGoogle Scholar
- Soerjomataram I, Louwman MW, Ribot JG, Roukema JA, Coebergh JW. An overview of prognostic factors for long-term survivors of breast cancer. Breast Cancer Res Treat. 2008;107:309–30.PubMed CentralView ArticlePubMedGoogle Scholar
- Lyman GH, Giuliano AE, Somerfield MR, Benson AB 3rd, Bodurka DC, Burstein HJ, et al. American Society of Clinical Oncology guideline recommendations for sentinel lymph node biopsy in early-stage breast cancer. J Clin Oncol. 2005;23:7703–20.Google Scholar
- Gipponi M, Bassetti C, Canavese G, Catturich A, Di Somma C, Vecchio C, et al. Sentinel lymph node as a new marker for therapeutic planning in breast cancer patients. J Surg Oncol. 2004;85:102–11.View ArticlePubMedGoogle Scholar
- Krag DN, Anderson SJ, Julian TB, Brown AM, Harlow SP, Costantino JP, et al. Sentinel-lymph-node resection compared with conventional axillary-lymph-node dissection in clinically node-negative patients with breast cancer: overall survival findings from the NSABP B-32 randomised phase 3 trial. Lancet Oncol. 2010;11:927–33.PubMed CentralView ArticlePubMedGoogle Scholar
- Kamath VJ, Giuliano R, Dauway EL, Cantor A, Berman C, Ku NN, et al. Characteristics of the sentinel lymph node in breast cancer predict further involvement of higher-echelon nodes in the axilla: a study to evaluate the need for complete axillary lymph node dissection. Arch Surg. 2001;136:688–92.View ArticlePubMedGoogle Scholar
- Rahusen FD, Torrenga H, van Diest PJ, Pijpers R, van der Wall E, Licht J, et al. Predictive factors for metastatic involvement of nonsentinel nodes in patients with breast cancer. Arch Surg. 2001;136:1059–63.View ArticlePubMedGoogle Scholar
- Pal A, Provenzano E, Duffy SW, Pinder SE, Purushotham AD. A model for predicting non-sentinel lymph node metastatic disease when the sentinel lymph node is positive. Br J Surg. 2008;95:302–9.View ArticlePubMedGoogle Scholar
- Barranger E, Coutant C, Flahault A, Delpech Y, Darai E, Uzan S. An axilla scoring system to predict non-sentinel lymph node status in breast cancer patients with sentinel lymph node involvement. Breast Cancer Res Treat. 2005;91:113–9.View ArticlePubMedGoogle Scholar
- Van Zee KJ, Manasseh DM, Bevilacqua JL, Boolbol SK, Fey JV, Tan LK, et al. A nomogram for predicting the likelihood of additional nodal metastasis in breast cancer patients with a positive sentinel node biopsy. Ann Surg Oncol. 2003;10:1140–51.View ArticlePubMedGoogle Scholar
- Albertini JJ, Lyman GH, Cox C, Yeatman T, Balducci L, Ku N, et al. Lymphatic mapping and sentinel node biopsy in the patient with breast cancer. JAMA. 1996;276:1818–22.View ArticlePubMedGoogle Scholar
- Kim T, Giuliano AE, Lyman GH. Lymphatic mapping and sentinel lymph node biopsy in early-stage breast carcinoma: a meta-analysis. Cancer. 2006;106:4–16.View ArticlePubMedGoogle Scholar
- Barone JE, Tucker JB, Perez JM, Odom SR, Ghevariya V. Evidence-based medicine applied to sentinel lymph node biopsy in patients with breast cancer. Am Surg. 2005;71(1):66–70.PubMedGoogle Scholar
- Kelley MC, Hansen N, McMasters KM. Lymphatic mapping and sentinel lymphadenectomy for breast cancer. Am J Surg. 2004;188(1):49-61.Google Scholar
- Giuliano AE, Jones RC, Brennan M, Statman R. Sentinel lymphadenectomy in breast cancer. J Clin Oncol. 1997;15(6):2345–50.PubMedGoogle Scholar
- Guenther JM, Krishnamoorthy M, Tan LR. Sentinel lymphadenectomy for breast cancer in a community managed care setting. Cancer J Sci Am. 1997;3:336–40.PubMedGoogle Scholar
- Simmons R, Thevarajah S, Brennan MB, Christos P, Osborne M. Methylene blue dye as an alternative to isosulfan blue dye for sentinel lymph node localization. Ann Surg Oncol. 2003;10:242–7.View ArticlePubMedGoogle Scholar
- Abe H, Mori T, Umeda T, Tanaka M, Kawai Y, Shimizu T, et al. Indocyanine green fluorescence imaging system for sentinel lymph node biopsies in early breast cancer patients. Surg Today. 2011;41(2):197–202.View ArticlePubMedGoogle Scholar
- Sugie T, Sawada T, Tagaya N, Kinoshita T, Yamagami K, Suwa H, et al. Comparison of the indocyanine green fluorescence and blue dye methods in detection of sentinel lymph nodes in early-stage breast cancer. Ann Surg Oncol. 2013;20(7):2213–8.View ArticlePubMedGoogle Scholar
- Polom K, Murawa D, Nowaczyk P, Rho YS, Murawa P. Breast cancer sentinel lymph node mapping using near infrared guided indocyanine green and indocyanine green–human serum albumin in comparison with gamma emitting radioactive colloid tracer. Eur J Surg Oncol. 2012;38:137–42.View ArticlePubMedGoogle Scholar
- Shiozawa M, Lefor AT, Hozumi Y, Kurihara K, Sata N, Yasuda Y, et al. Sentinel lymph node biopsy in patients with breast cancer using superparamagnetic iron oxide and a magnetometer. Breast Cancer. 2013;20:223–9.View ArticlePubMedGoogle Scholar
- Douek MKJ, Monnypenny I, Kothari A, Kothari A, Zechmeister K, Brown D, et al. Sentinel node biopsy using a magnetic tracer versus standard technique: the SentiMAG multicentre trial. Ann Surg Oncol. 2014;21:1237–45.View ArticlePubMedGoogle Scholar
- Sever A, Jones S, Cox K, Weeks J, Mills P, Jones P. Preoperative localization of sentinel lymph nodes using intradermal microbubbles and contrast-enhanced ultrasonography in patients with breast cancer. Br J Surg. 2009;96:1295–9.View ArticlePubMedGoogle Scholar
- Sever AR, Mills P, Jones SE, Cox K, Weeks J, Fish D, et al. Preoperative sentinel node identification with ultrasound using microbubbles in patients with breast cancer. AJR Am J Roentgenol. 2011;196:251–6.View ArticlePubMedGoogle Scholar
- Cox K, Sever A, Jones S, Weeks J, Mills P, Devalia H, et al. Validation of a technique using microbubbles and contrast enhanced ultrasound (CEUS) to biopsy sentinel lymph nodes (SLN) in pre-operative breast cancer patients with a normal grey-scale axillary ultrasound. Eur J Surg Oncol. 2013;39:760–5.View ArticlePubMedGoogle Scholar
- Sever AR, Mills P, Hyvelin JM, Weeks J, Gumus H, Fish D, et al. Percutaneous removal of sentinel lymph nodes in a swine model using a breast lesion excision system and contrast-enhanced ultrasound. Eur Radiol. 2012;22:545–50.View ArticlePubMedGoogle Scholar
- Kusnetzky L, Khalid A, Khumri T, Moe TG, Jones PG, Main ML. Acute mortality in hospitalized patients undergoing echocardiography with and without an ultrasound contrast agent: results in 18671 consecutive studies. J Am Coll Cardiol. 2008;51:1704–6.View ArticlePubMedGoogle Scholar
- Dijkmans P, Visser C, Kamp O. Adverse reactions to ultrasound contrast agents: is the risk worth the benefit? Eur J Echocardiogr. 2005;6:363–6.View ArticlePubMedGoogle Scholar
- Giuliano AE, Hunt KK, Ballman KV, Beitsch PD, Whitworth PW, Blumencranz PW, et al. Axillary dissection vs no axillary dissection in women with invasive breast cancer and sentinel node metastasis: a randomized clinical trial. JAMA. 2011;305(6):569-75.View ArticlePubMedGoogle Scholar
- Galimberti V, Cole BF, Zurrida S, Viale G, Luini A, Veronesi P, et al. Axillary dissection versus no axillary dissection in patients with sentinel-node micrometastases (IBCSG 23–01): a phase 3 randomised controlled trial. Lancet Oncol. 2013;14:297–305.PubMed CentralView ArticlePubMedGoogle Scholar
- Goldberg BB, Merton DA, Liu JB, Thakur M, Murphy GF, Needleman L, et al. Sentinel lymph nodes in a swine model with melanoma: contrast-enhanced lymphatic US. Radiology. 2004;230:727–34.View ArticlePubMedGoogle Scholar
- Omoto K, Matsunaga H, Take N, Hozumi Y, Takehara M, Omoto Y, et al. Sentinel node detection method using contrast-enhanced ultrasonography with sonazoid in breast cancer: preliminary clinical study. Ultrasound Med Biol. 2009;35:1249–56.View ArticlePubMedGoogle Scholar
- Harlow SP, Krag DN, Julian TB, Ashikaga T, Weaver DL, Feldman SA, et al. Prerandomization surgical training for the NSABP-B32 trial—a randomization phase III clinical trial to compare sentinel node resection to conventional axillary dissection in clinically node negative breast cancer. Ann Surg. 2005;241:48–54.PubMed CentralPubMedGoogle Scholar
- Veronesi, U. The sentinel node and breast cancer. Br J Surg. 1999;86:1-2.Google Scholar
- Goldhirsch A, Winer EP, Coates AS,Gelber RD, Piccart-Gebhart M, Thürlimann B, et al. Panel members. Personalizing the treatment of women with early breast cancer: highlights of the St Gallen international expert consensus on the primary therapy of early breast cancer 2013. Ann Oncol. 2013;24:2206–23.PubMed CentralView ArticlePubMedGoogle Scholar