Skip to main content

Table 2 Univariate and multivariate linear regression results of the mean glandular dose per view

From: Radiation dose affected by mammographic composition and breast size: first application of a radiation dose management system for full-field digital mammography in Korean women

 

Univariate linear regression

Multivariate linear regression

Beta (SE)

p value

R 2

Beta (SE)

p value

Adjusted R 2

Age (years)

−0.013 (0.001)

<0.0001

0.038

−0.002 (0.001)

0.0358

0.820

Mammographic breast size (mm)

0.004 (0.001)

<0.0001

0.016

0.002 (0.0004)

<0.0001

Mammographic breast composition

 Non-dense

Reference

 

0.094

Reference

 

 Dense

0.462 (0.031)

<0.0001

0.201 (0.017)

<0.0001

Breast-conserving surgery

 No

Reference

  

Reference

 

 Yes

0.249 (0.046)

<0.0001

0.014

0.038 (0.020)

0.0605

Peak kilovoltage (kVp)

0.074 (0.008)

<0.0001

0.039

  

Current (mAs)

0.014 (0.0004)

<0.0001

0.384

0.014 (0.0002)

<0.0001

Compressed breast thickness (mm)

0.013 (0.001)

<0.0001

0.053

0.008 (0.001)

<0.0001

Breast compression force (N)

0.006 (0.001)

<0.0001

0.058

0.003 (0.0003)

<0.0001

  1. Beta represents the unstandardized regression coefficient. Due to the high multicollinearity of kVp with several thicknesses, we removed kVp from the multivariate linear regression analysis