Skip to main content

Table 3 Breast CNB accuracy according to gauge, lesion type, lesion size and calcifications

From: Accuracy of 16/18G core needle biopsy for ultrasound-visible breast lesions

 

False-negative

Underestimation

Complete sensitivity

Absolute sensitivity

Agreement

 

n (%)

n (%)

%

n (%)

n (%)

18G (n = 389)

     

 Mass

4/327 (1.2)

19/38 (50)

98.8

308/327 (94.2)

343/366 (93.7)

 Non-mass

1/19 (5.3)

4/9 (44.4)

94.7

17/19 (89.5)

18/23 (78.3)

Mass Size (mm)

     

 ≤10

2/47 (4.3)

6/18 (33.3)

95.7

39/47 (83.0)

52/60 (86.7)

 11–20

1/134 (0.7)

7/10 (70)

99.3

129/134 (96.3)

141/149 (94.6)

 21–50

1/137 (0.7)

6/10 (60)

99.3

131/137 (95.6)

139/146 (95.2)

 >50

0/9 (0)

0/0 (0)

100

9/9 (100)

11/11 (100)

Calcifications

     

 Present

2/133 (1.5)

6/15 (40)

98.5

129/133 (97.0)

135/143 (94.4)

 Absent

3/213 (1.4)

17/32 (53)

98.6

196/213 (92.0)

226/246 (91.9)

Overall

5/346 (1.4)

23/47 (48.9)

98.6

325/346 (93.9)

361/389 (92.8)

16G (n = 566)

     

 Mass

7/488 (1.4)

33/68 (48.5)

98.6

460/488 (94.3)

505/545 (92.7)

 Non-mass

0/17 (0)

3/8 (37.5)

100

14/17 (82.4)

18/21 (85.7)

Mass Size (mm)

     

 ≤10

2/58 (3.4)

10/19 (52.6)

96.6

50/58 (86.2)

60/72 (83.3)

 11–20

3/208 (1.4)

14/24 (58.3)

98.6

198/208 (95.2)

215/232 (92.7)

 21–50

2/213 (0.9)

9/24 (37.5)

99.1

203/213 (95.3)

219/230 (95.2)

 >50

0/9 (0)

0/1 (0)

100

9/9 (100)

11/11 (100)

Calcifications

     

 Present

2/166 (1.2)

6/18 (33.3)

98.8

162/166 (97.6)

168/176 (95.5)

 Absent

5/339 (1.5)

30/58 (51.7)

98.5

312/339 (92.0)

355/390 (91.0)

Overall

7/505 (1.4)

36/76 (47.4)

98.6

474/505 (93.9)

523/566 (92.4)