From: An in-vitro animal experiment on metal implants’ thermal effect on radiofrequency ablation
Distance from RF electrode (cm) | Case number | Therm ometry site ( n) | Temperature of test group (, °C) | Temperature of control group (, °C) | Paired- samples t test | ||||
---|---|---|---|---|---|---|---|---|---|
Ablation 8 min | Ablation 12 min | Ablation 15 min | Ablation 8 min | Ablation 12 min | Ablation 15 min | ||||
1.0 | 14 | 42 | 71.7±6.79 | 78.0±4.81 | 81.3±5.32 | 71.1±7.84 | 78.9±5.13 | 80.9±5.02 | P>0.05 |
2.0 | 14 | 42 | 43.3±2.55 | 43.3±1.84 | 46.1±3.50 | 44.2±3.89 | 44.2±3.20 | 47.1±2.83 | P>0.05 |
3.0 | 14 | 42 | 29.8±2.16 | 33.1±2.58 | 36.5±2.48 | 30.5±1.98 | 33.7±2.32 | 36.8±2.01 | P>0.05 |